Model 200/1 Hydrocarbon Analyzer -
FID Methane / Non Methane (NMHC), Oven Heated Hydrocarbon Analyzer

The VIG Industries, Inc. Model 200/1 is a microprocessor based, oven heated methane / non methane (NMHC) hydrocarbon analyzer for high accuracy, sensitivity and stability. The Model 200/1 measures methane and non methane hydrocarbons at preset intervals. The VIG Model 200/1 uses one(1) Flame Ionization Detector (FID) to obtain methane and non-methane hydrocarbon readings.
A sample is fed to the analyzer via an internal heated pump. After the sample reaches the analyzer, a portion of the sample is trapped and fed
through a GC (gas chromatography) column to separate the methane from the total and is fed to the FID detector for analysis.
As soon as the methane is read, the remaining sample is back flushed through the GC column and fed to the FID detector for analysis of the non-methane component. The LCD display indicates the amount of methane and non-methane hydrocarbons.
The Model 200/1 utilizes one flame ionization detector (FID). The detector, all sample lines, GC column and all solenoid valves that come in contact with the sample are maintained in a controlled, heated oven to prevent condensation, and to provide reliable performance in the analysis of concentrations of wide varieties of hydrocarbons in gaseous mixtures and in ambient air.
Model 200/1 Brochure (1.3MB)
Applications
-
Compliance Monitoring - US EPA Method 18
-
Process Monitoring - Continuous monitoring and alarm or control of:
-
Efficiency Monitoring - Monitoring effluent of volatile organic compound (VOC) reduction equipment for:
-
Environmental compliance
-
Efficiency control of incinerators (Thermal or catalytic)
-
Scrubbers
-
Carbon absorbers
-
Monitoring of catalytic converters
-
Other abatement equipment
-
Safety Monitoring - Lower explosive limit (LEL) monitoring and/or control of:
-
Ovens/dryers
-
Fugitive emissions monitoring
-
Personnel work area monitoring
-
Leak detection of process equipment or solvent storage areas
-
Stack Monitoring
Features
-
Easy to use software
-
GC technology used for separation
-
Automatic startup / ignition
-
Automatic fuel shut-off system
-
Automatic flame-out indicator
-
Adjustable alarm and oven settings
-
Precision 1% of full scale
-
Fast response
-
Two stage sample filter with exchangeable sintered stainless steel elements
-
Teflon® isolated detector (FID)
-
19" rack or bench mount
-
Heated sample pump head
-
Outputs 0-1, 0-5, or 0-10Volts
-
Automatic ignition
Options
-
Zero and calibration input ports with software
-
4-20mA output - no extra charge
-
Range recognition relays
-
RS-232 interface
-
Internal combustion air supply
-
Automatic sample shut-off system
Related Available Equipment
-
External zero air generator (reduces bottles)
-
Hydrogen Generator (reduces bottles)
-
Heated sample lines and controllers
-
Strip chart recorders and data loggers
-
NEMA rated enclosures
Standard Specifications
| |
| Analysis Method |
Oven Heated, Flame Ionization Detector (FID) |
| Separation Method |
Column |
| Precision |
+/- 1% of Full Scale |
| Response Time |
Methane - Approximately 40 seconds,
updated every 3 minutes
Non methane - Approximately 70 seconds,
updated every 3 minutes |
| Zero Drift |
+/- 1% of Full Scale Per 24 Hours |
| Span Drift |
+/- 1% of Full Scale Per 24 Hours |
| Linearity |
Within 1% of full scale |
| Oxygen Synergism |
Within 1% of full scale within selected range |
| Ranges |
(4 Ranges per amplifier, two amplifiers per analyzer,
one amplifier for total hydrocarbons,
one amplifier for methane and non-methane)
Standard Ranges -
0-100, 1000, 10000, 100000ppm (LDL 0.1ppm)
0-10, 100, 1000, 10000ppm (LDL 0.01ppm) |
| Display |
Graphic, backlit, 240W x 64H pixels, high contrast,
wide viewing angle |
| Outputs |
0-1, 5, or 10 Volts & optional 4-20mA |
| Alarms |
Flame Out & Concentration, Latching |
| Safety |
Flame Out Indicator with Fuel Shut-Off System |
| Oven Temperature |
Operating Temperature 275°F, Adjustable 200 to 300° F |
| Sample Pump |
Heated 4 Liters/Minute standard or Heated 14 Liters/Minute optional |
| Ignition |
Automatic (can be set to manual by operator) |
| Glow Plug |
Main and spare glow plug, selectable by switch on back panel |
| Dimensions |
19" front panel, 16.75" wide chassis, 18" deep chassis,
21" deep with fittings and handles, 9" high |
| Weight |
55 to 65 pounds depending on options |
|
Operating Requirements
| |
| Fuel |
Hydrogen (UHP Grade) @ 18psi incoming pressure |
| Combustion Air |
Zero Grade Air @ 18PSI |
| Operation Air |
Compressed Air (Instrument Air - Dry / Oil-Less) @ 60PSI |
| Zero Gas |
Zero Air (UHP Grade) @ 10PSI |
| Calibration Gas |
User Selected @ 10PSI |
| Carrier Gas |
Nitrogen (UHP Grade) @ 30PSI |
| Power |
115Vac, 60Hz or optional 220Vac, 50Hz |
| Ambient Temperature |
32 to 110º F |
|
Warranty
All instruments sold by VIG Industries, Inc., are warranted for a period of one (1) year from the date of purchase, against defects in materials and workmanship. VIG warrants that the product supplied conforms to the specifications assigned thereto. There is no other warranty either implied or expressed. Seller liability is limited specifically to the cost or assigned value of the items sold.
|
|